Our Products
 
 
Our Principles



 
 
 
 
 
Triple Micro Raman Spectrometer
 
 
Triple Micro Raman Spectrometer
STR-555/557/777
   
Low Frequency Raman Bands are Measurable as Close as 5cm-1 to the Laser Line
   
 
The stray light rejection of the triple subtractive mode enables low frequency measurements down to as close as 5cm-1 to the Raleigh line. The resolution of the triple additive mode allows for spectral resolutions of 0.008nm (0.22cm-1) @ 600nm with three 1800 g/mm gratings. Even higher resolutions are attainable using higher groove density gratings. The STR-777 has the highest resolution of any commercially available triple spectrometer in both additive and subtractive mode.
 
Subtractive mode:- : Excellent stray light rejection with a
CCD detector can be achieved when subtractive mode is
used. (right figure)
 
Additive mode:- the gratings on all 3 stages contribute to positive light dispersion. (left figure)
 
   
 
Additive mode high resolution data   Subtractive mode low frequency data
 
STR-555/557/777 Specification of Triple Raman System
Laser
HeCd-X/Y DL532-50, 100 Ar-50,100,2W HeNe-17, 35 DL785-50,100,300
He Cd Laser
X:325nm, 10,20,35,50mW Y:442nm, 40,55,75mW
Diode Green Laser
532nm
50, 100mW
Air Coolde Ar+    Laser
488nm: B, 514nm: G
50, 100mW, 2W
He-Ne Laser
633nm
17, 35mW
Diode Near-IR Laser
785nm
50,100,300mW
Standard items with the laser are 3m optical fiber and laser to fiber coupler. Please consult us for additional specification. Option: SDT: Step down transformer 440VAC -> 220VAC, 3 phase, 50/60Hz
Imaging Spectrograph
STR555 STR557 STR777
1st, 2nd,Stage Additive, Subtractive Mode: 500mm FL, f/5.9 Additive, Subtractive Mode: 500mm FL, f/5.9 Additive, Subtractive Mode: 750mm FL, f/9.8
3rd    Stage 500mm FL, f/5.9 750mm FL, f/9.8 750mm FL, f/9.8
Resolution* Subtractive: 0.59cm-1          Additive: 0.21cm-1 Subtractive: 0.39cm-1          Additive: 0.17cm-1 Subtractive: 0.39cm-1          Additive: 0.13cm-1
Low frequency 5cm-1               (2cm-1  at PMT ) 5cm-1               (2cm-1  at PMT ) 5cm-1 (2cm-1    at PMT )

Common items to the above are aberration corrected Czemy-Turner single spectrograph, 3 gratings (max. 9 gratings) and a Windows based computer with a date collection/processing software. Entrance slit : 10um ~ 3.0mm, RS232C/USB, optical fiber 3m.
Spectrum Rangel: 200~2000nm (depend on grating),   Raman shift : 5 ~ 5000cm-1 L: w/o Grams 32/AITM     data processing software   * 1800g @500nm

Cooled CCD camera

 

100/400-F,B,B_eXcelon,BR 256E DU401/420A-UV,OE,BV,BR-DD 2KF,2KB(UV),2KB eXcelon
format 1340       X     100    (400)     pixels,
20x20μm
1024 X 256 pixels, 26x26μm 401:1024 X 128 pixels, 26x26μm
420:1024 X 256 pixels, 26x26μm
2048 X 512 pixels, 13.5x13.5μm
QE F: 45%    B,B_eXcelon: 95% BR: 90% E : 60% UV: 45%         OE: 55% BV, BR-DD: 90%

2KF: 40%
2KB,2KB eXcelon: 95%

Dark noise
(e-/pixel/sec)
 0.0025 typ for 100B,  B_eXcelon
0.02 typ   for 100BR at -80℃
0.003 typ for 256E at -75℃ 0.002 typ for UV at -70℃
0.005 typ for BV at -70℃
0.003 typ for BR-DD at –90℃
0.001 typ at -75℃
Read noise 3 e- rms typ at 100KHz 6 e- rms typ at 100KHz 5 e- rms typ at 50KHz 3.5 e- rms typ at 100KHz
TE cooling -75℃ :400, -80℃: 100 -75℃ DU:-70 ℃(-85 ℃),  DU-BR-DD:-80 ℃
(-100℃)
-75℃
CCD type F/UV: front illuminated(FI) /w UV coat, B/BV:back illuminated  (BI, BI_eXcelon), BR-DD:BI deep depletion(DD), E: open electrode
ADC 16 bit, USB, wavelength range: 200-1050nm /w UV coat option, , TE cooling temperature: (   ) option
TE-PMT: Thermal electric cooling PMT detector (single channel) with housing, power supply, data collection unit

Raman sampling accessories

Optical microscope   ST-BX51*

Remote Raman Probe   RPM-XX

 

Confocal Raman optical microscope with < 1μm spatial resolution using x100 objective lens
Raman probe with Raman filter set   (1 wavelength)*2    , halogen light (ref/trans),
Objective lens : x10, x20LWD (WD=25mm), x40X(Near UV), x50, x100, CCD color video camera

 

>25mm working distance, spot size < 5um
Raman filter set   (1 wavelength)*2

Raman filter set is common for both optical microscope and remote Raman probe and is easily exchangeable to other laser lines.
*M option: motorized revolver and halogen illuminator   *P option: Polarized Raman measurement, and observation    *2    Please specify the excitation wavelength:

325nm ~150(30) cm-1, 488, 514.5, 532nm, ~50(10) cm-1, 633nm ~40(5)cm-1, 785nm ~30(5)cm-1 ( ) Option,   Please consult us for additional specifications